Reversed-Phase HPLC Analysis of Levetiracetam in Tablets Using Monolithic and Conventional C18 Silica Columns
Abstract
Development and validation of an RP-HPLC method for determination of levetiracetam in pharmaceutical tablets is described. The separation and quantification of levetiracetam and caffeine (internal standard) were performed using a single analytical procedure with two different types of stationary phases, conventional Phenomenex Gemini C18 (100 x 4.6 mm, 5 mu m) and Merck Chromolith Performance RP18e (100 x 4.6 mm, macropore size 2 mm, micropore size 13 nm) monolithic silica. Five-microliter aliquots of samples were injected into the system and eluted using water acetonitrile (90 + 10, v/v) mobile phase pumped at the rate of 1 mL/min. The analyte peaks were detected at 200 nm using a diode array detector with adequate resolution. Validation studies were performed using the method recommended by the International Conference on Harmonization, the U.S. Pharmacopeia, and AOAC INTERNATIONAL, which includes accuracy, precision, range, limits, robustness, and system suitability parameters. Levetiracetam and caffeine were detected in about 7 min using the conventional column, whereas less than 5 min was required when the monolithic column was used. Calibration plots had r values close to unity in the range of 0.8-8.0 mu g/mL. Assay of levetiracetam in a tablet formulation was demonstrated as an application to real samples.
Source
Journal of Aoac InternationalVolume
93Issue
4Collections
- Makale Koleksiyonu [115]
- PubMed İndeksli Yayınlar Koleksiyonu [1498]
- WoS İndeksli Yayınlar Koleksiyonu [7605]